Introduction
|
|
The FMS 1 modulated chlorophyll fluorometer is a versatile pulse modulated
instrument with mains power supply for laboratory/near field use. It is suitable
for non-invasive sampling of chlorophyll fluorescence under natural or
artificial light conditions & may be interfaced with our range of oxygen
electrode chambers for simultaneous oxygen measurements.
The FMS 1 modulated chlorophyll fluorometer consists of a control unit housing
all of the electronics, optics and light sources necessary to derive most common
chlorophyll fluorescence parameters. These are optically linked to the sample by
a statistically randomised fibre optic cable that is suitable for insertion into
a range of sample containers such as oxygen electrodes, gas analysis chambers,
petri dishes and microtitre plates.
The system may be operated in several different modes: serial connection to a
Windows® PC enables real-time instrument control and data presentation.
|
|
|
Captured data is simultaneously presented as a real-time chart recorder
emulation and parameters-only format for easy identification of key experimental
events. This PC mode of operation is suitable for development of complex
protocols which may be programmed into the instrument using the simple drag and
drop editor to generate user-defined scripts. These scripts automate the
execution of experiments, allowing complex protocols involving many control
events to be operated with the same ease as single control event measurement
such as Fv/Fm. Once programmed, the FMS 1 chlorophyll fluorometer can be
operated as a stand-alone fluorimeter inside the laboratory or outside (via
connection to an optional external battery). All measurement data and calculated
parameters are saved to integral protected memory. The unit can store up to six
experimental protocols, any one of which may be accessed and executed using the
built-in menu system. When data collection is complete the results can be
downloaded to the Windows® software for full analysis.
All of the light sources required for modulated measurement of common
chlorophyll fluorescence parameters are self-contained within the instrument.
| |
594 nm amber modulating beam with 4 step frequency control. (Optional 470 nm
blue LED)
|
| |
Dual-purpose halogen light source providing actinic light (0 - 3000 µmol m-2
s-1 in 50 steps) and saturating pulse (0 - 20,000 µmol m-2
s-1 in 100 steps).
|
| |
735 nm far-red LED source for preferential PSI excitation allowing accurate
determination of the Fo' parameter.
|
|