You are in: Products >
Chlorophyll Fluorescence >
Continuous Excitation
> Handy PEA Introduction
Introduction
|
|
The Handy PEA chlorophyll fluorimeter consists of a compact, light-weight
control unit encapsulating sophisticated electronics providing the high time
resolution essential in performing measurements of fast chlorophyll fluorescence
induction kinetics.
The chlorophyll fluorescence signal received by the sensor head during recording
is digitised within the Handy PEA control unit using a fast Analogue/Digital
converter. The chlorophyll fluorescence signal is digitised at different rates
dependent upon the different phases of the induction kinetic. Initially, data is
sampled at 10µs intervals for the first 300µseconds.
|
|
|
Simple to configure and operate, the Handy PEA chlorophyll fluorimeter features
the capacity to store up to 5 user-defined protocols for different field
applications. Protocols are written using a custom Windows® software package,
PEA Plus (supplied). This allows single or multiple measurement assays with
optional pre-illumination periods to be defined and uploaded to the memory of
Handy PEA via RS232 serial communications. The use of protocols ensures maximum
reproducibility of results during field applications involving large scale
screening away from a laboratory environment. A waterproof, tactile keypad
allows selections and inputs to be made and a liquid crystal display module
presents menu options and data.
Up to 1000 recordings of between 0.1 - 300 seconds may be saved in the memory of
Handy PEA chlorophyll fluorimeter. Saved data may be viewed onscreen in
numerical format with calculated parameters or transferred to the PEA Plus
software where it may be viewed graphically or exported to external software
packages for further statistical analysis.
The sensor unit consists of an array of 3 ultra-bright red LED’s optically
filtered to a peak wavelength of 650 nm, which is readily absorbed by the
chloroplasts of the leaf, at a maximum intensity of >3000 µmol m-2 s-1
at the sample surface. The LED's are focused via lenses onto the leaf surface to
provide even illumination over the area of leaf exposed by the leafclip (4mm
dia). LED's have the advantage of being rugged, emitting low levels of heat, and
of rising to full intensity very rapidly (typically microseconds) after being
switched on. This feature eliminates the inaccuracies of Fo measurement and the
constraints on speed and reliability associated with a shutter which is a
necessary item in systems using filament lamps rather than LED's.
An optical feedback circuit monitors and corrects changes in the output
intensity of the LED's. These changes are caused by internal heat build up in
the LED's. The circuit also compensates for intensity changes caused by
variation in ambient temperature. The light source is calibrated before leaving
the factory but may be calibrated by the user at regular intervals using the SQS
Serial Quantum Sensor.
The detector is a high performance Pin photodiode and associated amplifier
circuit. The optical design and filtering ensure that it responds maximally to
the longer wavelength fluorescence signal and blocks the reflected shorter
wavelength LED light used as the source of illumination.
The sensor unit is connected to the Handy PEA control unit via a standard
connection cable of 1m in length however, connection cables of up to 10m in
length are also available on request.
|