Handy PEA Continuous Excitation Chlorophyll Fluorimeter

Introduction

The Handy PEA chlorophyll fluorimeter consists of a compact, light-weight control unit encapsulating sophisticated electronics providing the high time resolution essential in performing measurements of fast chlorophyll fluorescence induction kinetics.

The chlorophyll fluorescence signal received by the sensor head during recording is digitised within the Handy PEA control unit using a fast Analogue/Digital converter. The chlorophyll fluorescence signal is digitised at different rates dependent upon the different phases of the induction kinetic. Initially, data is sampled at 10µs intervals for the first 300µseconds.

Handy PEA Continuous Excitation Chlorophyll Fluorimeter

Simple to configure and operate, the Handy PEA chlorophyll fluorimeter features the capacity to store up to 5 user-defined protocols for different field applications. Protocols are written using a custom Windows® software package, PEA Plus (supplied). This allows single or multiple measurement assays with optional pre-illumination periods to be defined and uploaded to the memory of Handy PEA via RS232 serial communications. The use of protocols ensures maximum reproducibility of results during field applications involving large scale screening away from a laboratory environment. A waterproof, tactile keypad allows selections and inputs to be made and a liquid crystal display module presents menu options and data.

Up to 1000 recordings of between 0.1 - 300 seconds may be saved in the memory of Handy PEA chlorophyll fluorimeter. Saved data may be viewed onscreen in numerical format with calculated parameters or transferred to the PEA Plus software where it may be viewed graphically or exported to external software packages for further statistical analysis.

The sensor unit consists of an array of 3 ultra-bright red LED’s optically filtered to a peak wavelength of 650 nm, which is readily absorbed by the chloroplasts of the leaf, at a maximum intensity of >3000 µmol m-2 s-1 at the sample surface. The LED's are focused via lenses onto the leaf surface to provide even illumination over the area of leaf exposed by the leafclip (4mm dia). LED's have the advantage of being rugged, emitting low levels of heat, and of rising to full intensity very rapidly (typically microseconds) after being switched on. This feature eliminates the inaccuracies of Fo measurement and the constraints on speed and reliability associated with a shutter which is a necessary item in systems using filament lamps rather than LED's.

An optical feedback circuit monitors and corrects changes in the output intensity of the LED's. These changes are caused by internal heat build up in the LED's. The circuit also compensates for intensity changes caused by variation in ambient temperature. The light source is calibrated before leaving the factory but may be calibrated by the user at regular intervals using the SQS Serial Quantum Sensor.

The detector is a high performance Pin photodiode and associated amplifier circuit. The optical design and filtering ensure that it responds maximally to the longer wavelength fluorescence signal and blocks the reflected shorter wavelength LED light used as the source of illumination.

The sensor unit is connected to the Handy PEA control unit via a standard connection cable of 1m in length however, connection cables of up to 10m in length are also available on request.
 